Class-II Type B1 Biological Safety Cabinets / Bio Safety Cabinets
Bio safety cabinets are available in Stainless steel
and GI with PU coated construction. Worktables are of removable
type and suitable for oven baking. Inner surface are completely
line with stainless steel. HEPA filters are of Minipleat type.
Differential pressure is monitored through specially designed low-pressure
gauges. Working chamber is provided with UV & Illumination tubes.
Feather-touch control switch with indicators are provided for the
operations of Blower / Illumination / UV - Ultra Violet Lamp
Biosafety cabinets are available with various optional Special features
such as VFD controlled air-blowers for adjustable airflow, Operation
of UV - Ultra Violet Lamp integrated with the front-door movement to avoid accidental
exposure to UV - Ultra Violet rays, warning alarm when differential pressure exceeds
the limit to indicate filter choking level, warning alarm when the
front sash is opened beyond safety-level
Biosafety cabinets are available in various standard sizes and custom-built
sizes to suit user's requirement
Bio Hazard Safety Cabinet : Class II Type B1
Recommended for Product, Personnel & Environmental protection
In Class II Type B1 Biosafety Cabinets 30% of air re-circulated
within the cabinet and the remainder exhausted outside the laboratory
through optional exhaust air incinerator for environmental protection.
Velocity of airflow to the work zone creates an ultra-clean environment
for product protection.
Maintains a minimum measured average inflow velocity of 75 fpm through
the work area access opening. The face velocity at the access opening
varies between 80 fpm and 100 fpm. Have HEPA filtered down flow
air from a common plenum.
| Air Cleanliness |
Class 5 air cleanliness as per ISO 14644-1 ,
Class 100 air cleanliness as per US Federal Standard 209E |
| Air Balancing |
30% re-circulation & 70% exhaust |
| Particle retention |
0.3 Micron and above |
Note - Optionally ULPA filters can also be fitted to
acheive ISO Class 3 air cleanliness within workzone as per ISO 14644-1
(Equivalent to Class 1 air cleanliness as per US Federal Standard
209E, 100 times better than the usual class 100 air cleanliness
classification)
|
|