   Frequently Asked Questions in Cleanroom Construction and Maintenance   Is DOP aerosol harmful to the filters ?

No! Not if used properly. Tests were run on a HEPA filter backed up with charcoal filter, exposing the unit to 100 micrograms per litre of DOP for two hours. NO DOP penetration could be detected. Clean air was blown through the system for another hour and then the charcoal was analyzed for traces of DOP. The result was zero.