Frequently Asked Question - Biological Safety
|
| |
Comparison
of Biological Safety Cabinets
|
Type
|
Face velocity (lfpm)
|
Airflow Pattern
|
Radionuclides/
Toxic Chemicals
|
Biosafety Level(s)
|
Product Protection
|
| Class I* open front |
75
|
In at front; rear and top through HEPA filter |
No
|
2,3
|
No
|
| Class II Type A |
75
|
70% recirculated through HEPA; exhaust through HEPA |
No
|
2,3
|
Yes
|
| Type B1 |
100
|
30% recirculated through HEPA; exhaust via HEPA
and hard ducted |
Yes (Low levels
or volatility)
|
2,3
|
Yes
|
| Type B2 |
100
|
No recirculation; total exhaust via HEPA and hard
ducted |
Yes
|
2,3
|
Yes
|
| Type B3 |
100
|
Same as IIA, but plena under negative pressure to
room and exhaust air is ducted |
Yes
|
2,3
|
Yes
|
| Class III |
NA
|
Supply air inlets and exhaust through 2 HEPA filters |
Yes
|
3,4
|
Yes
|
* Glove panels may be added and will increase
face velocity to 150 lfpm; gloves may be added with an inlet
air pressure release that will allow work with chemicals/radionuclides.
|
|
|
| |
|
|